Search

PCMT PLUS Constant Temperature Oscillator

PCMT PLUS Constant Temperature Oscillator

The PCMT PLUS provides powerful mixing and temperature control of samples in microtubes or PCR plates. This model differs from the PCMT and PHMT in that it is equipped with a refrigeration function and at the same time enables control of seven devices from a PC via a Bluetooth connection. Software features include control of speed, temperature, time, alarms, program analysis, visualization of temperature vs. time and speed vs. time graphs, data export to Excel and CSV formats, and error messages/troubleshooting.

 

The heating and mixing functions can run simultaneously or independently. A total of five heating and cooling modules are available, including one with a plastic lid for PCR plates. All modules are interchangeable and easy to install.

Oscillation rate: 250 ~ 1400 rpm

 

Temperature setting range: +4 ~ 100 °C

Temperature control range: 15 ~ 100 °C above room temperature

Clear LCD display.

Displays each value of the setting and actual operation

Small footprint.

Saves bench space

Bluetooth connectivity enables control of 7 devices from a PC

 

The heating and mixing functions can run simultaneously.

It can also be run independently.

digital timer.

With audible alarm

Five heating and cooling blocks available

 

Product Model

PCMT PLUS Constant Temperature Oscillator

Setting range

4 °C to 100 °C

Control Range

15 °C to 100 °C at room temperature

Setting Resolution

0.1 °C

Stabilization (37 °C)

±0.1 °C

Accuracy (37 °C)

±0.5 °C

Platform Uniformity

±0.6 °C @4 °C / ±0.1 °C @37 °C / ±0.3 °C @100 °C

Average heating speed

5 °С/minute from 25 °С to 100 °С)

Average cooling speed

5 °С/min (from 100 °C to 25 °C)

1.8 °С/min (from 25 °C to 4 °C)

Calibration Options

 

be

Calibration Coefficient Range

0.936 to 1.063 (± 0.063)

Speed range

250-1400 rpm

RPM Setting Resolution

10 rpm

Maximum speed deviation

2 % @250 rpm

0.7 % @1400 rpm

Track Diameter

2 mm

 

timer

1 minute to 96 hours

Time Setting Resolution

One minute.

Maximum Continuous Operation Time

168 hours.

Display Screen

2-line 16-bit LCD display

sizes

220x240x130 mm

Input Voltage and Current/Power Consumption

12 V, 4.9 A / 60 W

External Power Supply

Input AC 100-240 V, 50/60 Hz, Output DC 12 V

weights

3.7 kg

 

Genetic analysis - extraction of DNA, RNA and further sample preparation

Biochemical studies of enzymatic reactions and processes

Extraction of metabolites from cells

Contact Hongke to help you solve your problems.