PCMT Coolable Constant Temperature Oscillator

PCMT Coolable Constant Temperature Oscillator

The PCMT Coolable Constant Temperature Oscillator is capable of both heating and cooling, and is adjustable in speed and temperature as needed. Suitable for microtubes and PCR plates, it is a versatile 3-in-1 device that can be used in combination with temperature control (heating/cooling) and mixing functions or independently.

Oscillation rate: 250 ~ 1400 rpm

 

Temperature setting range: +4 ~ 100 °C

Temperature control range: 15 ~ 100 ℃ at room temperature.

Temperature calibration function.

Adjustment of temperature difference between different microtubes and samples

Combined or individual use of temperature control (heating/cooling)

and hybrid functions

Temperature stability: ±0.1 °C

 

Rapid heating and cooling.

Quickly achieves the required mixing speeds

digital timer.

With audible alarm

Choose from five interchangeable

and applicable heating blocks

Dimensions (H x D x W)130 x 230 x 205 mm
Temperature Range+4 to 100°C
Temperature control rangeBelow ambient temperature 15 to +100°C
Uniformity @4°C0.6±℃
Uniformity @ 37°C0.1±℃
Uniformity @100°C0.3±C
HC18 block rate from 100°C to 25°C (average temperature rise)5°C/minute (approx. 15 minutes)
HC18 Module speed from @25°C to 4°C (average temperature rise)1.8°C/minute (approx. 12 minutes)
Oscillation speed250 to 1400 rpm
Display (temperature)2-line x 16-character LCD
quantitative (science)View Accessories
quantitative (science)1 microplate
Track Diameter2 mm
Timer (with auto shut-off and audible alarm)1 minute to 96 hours (1 minute increments)
Maximum noise53.8 decibels
Heating/cooling power60W
External Power SupplyInput AC 120-230V, 50/60Hz Output DC 12V
energy consumption60 watts (4.9 amps)
Input Voltage12V DC
weights3.7 kg

 

Life Sciences

Genetic analysis, DNA/RNA extraction and further sample preparation, biochemical studies of enzymatic reactions and processes, extraction of metabolites from cellular material, sample incubation, mixing and digestion for peptide profile analysis

Contact Hongke to help you solve your problems.