T100 Series Heating Circulating Water Bath
The T100 series of heating circulators consists of the Optima™ T100 Heating Circulator and a variety of bath sizes and materials to provide precise temperature control for applications in the 0°C to 100°C range. It can also be combined with the T-CLAMP to make it suitable for any vertical side bath.
The Optima™ heating circulator has built-in adaptive intelligent PID temperature control with Optimisation™ (ICO) and an integrated liquid pump, making the Optima™ heating circulator suitable for both sample insertion and temperature control of circulating liquids in external equipment.
Application Areas:
- Clinical, Microbiology and Pathology Laboratory - Heating, melting of culture media & culture samples
- Higher Education Research - Temperature Control of Spectrophotometers, Refractometers and Jacketed Vessels
- Industrial laboratory - temperature probe calibration, water analysis, QC testing, petrochemical product testing, material testing and milk sample testing
Temperature Setting Range
0 ~ 100 °C
Stability. Up to ±0.05ºC
Simple to use, with only one dial knob and two
functionalityKey, Quick Temperature Setting and Navigation Menu
User calibration, at the required operating temperature
Achieve optimal accuracy
Visual alarms.
Alerts the user in case of abnormal conditions.
1 set temperature.
Adding 3 adjustable preset temperatures makes operation more convenient.
4-digit display.
Easy reading of data from a distance
Low level protection and over-temperature protection
- Technical Parameters
Dimensions (H x D x W) | 333 x 172 x 120 mm |
Stability (DIN 12876) @70°C | 0.05±℃ |
Uniformity (DIN 12876) @70°C | 0.1±℃ |
Setting Resolution | 0.1°C |
showcase | 4-digit LED |
Timer Function | - |
Temperature preset quantity | 3 |
Recalibration point | 2 |
External probe socket (TXPEP, TXSEP) | - |
communication interface | - |
programmable | - |
Relay | - |
Safety - Ultra-temperature | settled |
Level - Float | - |
Language Options | - |
Alarm (configurable as a switching relay) | - |
Heater Power Supply - 230V | 1290W |
Heater Power - 120V | 1440W |
Power - 230V | 1400 (50-60 Hz) |
Power - 120V | 1500 (50-60 Hz) |
Height above tank rim | 200 mm |
Depth below tank rim | 135 mm |
Maximum pressure (water) | - |
Maximum flow rate (water) | - |
Pump Connector 6mm Orifice | - |
Pump Connector 9mm Aperture | - |
weights | 2.1 kg |