Search

ELLAB Temperature Standard

ETS25

The traceable ETS enables extremely accurate temperature measurements by converting resistance to temperature using its built-in algorithm.

 

When connected to the ValSuite software, the tank and ETS provide a perfect automatic setup. This saves time and facilitates calibration.

 

The ETS has a built-in digital display that shows readings in real time and can be powered via USB or LAN (PoE) depending on the user's preference.

 

  • Temperature range: -196 to +420°C
  • Accuracy: ±0.025 °C
  • Separation of electronic equipment from the heating zone
  • Rugged and versatile design

Shell material:

Recyclable ABS plastic

IP rating of the housing:

41

Show:

High resolution 3-digit display

Interface and power supply:

USB or Power over Ethernet (PoE)

Temperature range:

-196 to +420°C

Accuracy - Includes instrumentation, PRT and drift:

-50 to +150°C:

±0.025 °C

Other temperatures:

±0.025 °C

Measure the sampling rate:

1 sec.

Digital Filter:

0-10 seconds

Resistance:

100 Ω

Sensor response time:

T-63%:

33 seconds.

T-90%:

45 seconds.

Sensor material:

Chrome-nickel-iron alloy™ 600

Sensor Size:

ø 6.35 x 300 mm

Minimum immersion depth:

100 mm

Environment:

5-50°C/0-90% Relative Humidity, Non-Condensing

Calibration:

Dates back to the Teknologisk Institute, Denmark.

Calibration algorithm:

ITS-90

Suggested Calibration Interval:

12 months

Weight:

305 g

Other Options

Key Benefits and Features

  • Temperature range: -80 to +250 °C
  • Accuracy: ±0.015 °C / ±0.025 °C
  • Separation of electronic equipment from the heating zone
  • Rugged and versatile design
  • USB and LAN communication options
  • Bracket can be easily mounted on wall or steel surface
  • Built-in digital display for real-time data readout

Contact Hongke to help you solve your problems.

Let's have a chat