Search

TrackSense Pro XL-剛性和智能柔性溫度傳感器

TrackSense Pro XL 無線數據記錄儀

剛性和智能柔性溫度傳感器

剛性和智能柔性溫度傳感器配備了用於難以到達的位置的柔性傳感器和不銹鋼彎曲傳感器(彎曲 90° 或 180°)。它的溫度範圍為 -80 °C 至 +140 °C,具有無與倫比的 ± 0.05 °C 精度(在 -25 °C 至 +140 °C 範圍內)。它是食品行業應用的理想選擇,例如熱滲透和熱分布研究。

 

TrackSense Pro XL 數據記錄儀系列配備了可互換傳感器選項,可解鎖多個測量參數:

 

✓ Temperature

✓ Pressure

✓ Conductivity

✓ Relative Humidity

✓ Vacuum

✓ Vapor penetration

 

The DataCORDER can be easily activated and read by multiple readout stations and combined with the Sky Module via wireless communication to obtain real-time data.

Sensors with this logger configuration:剛性和智慧柔性溫度傳感器Recorders with this sensor configuration:PRO XL 3G
Temperature measurement range:-80 至+140 °C / -196 至-80 °C,當記錄器保持在環境溫度時Working temperature:-80 至+140°C
Measurement Principle:ResistanceOperational pressure:0.001 mBar to 4 Bar ABS
Sensor element:Pt1000Surface material:316 Liter Stainless Steel
直徑:2 mm/ 1.8 mmDiameter of the logger:25mm
長度:10 mm/100 mmRecorder length:68.8mm
Measurement point from the tip position:3 mmWith battery weight:80g
Precision: Memory capacity:120000個數據點/60000個樣本值
-50 至-40°C:± 0.2 °C / ± 0.2 °CMinimum sampling rate:1s
-40 至-25°C:± 0.1 °C / ± 0.1 °CMaximum sampling rate:24h
-25 至+140°C:± 0.05 °C / ± 0.1 °CMaximum startup delay:14 days
Sensor response time: Benan:Ex II1GD Ex ia IIC T3 Ga,-55 °C ≤ Tamb ≤ +105 °C
T-63%:0.8/2.2 sTime accuracy:±5s/24h
T-90%:1.7/4.3 sBattery:TSP Standard Battery
  請注意:如果設備在ATEX 環境中使用,則必須考慮ATEX 憑證第17 節中規定的安全使用的特殊條件。


Sensor Options

Key Benefits and Features

  • One Logger - Multiple Sensor Options
  • Conforms to 21 CFR Part 11
  • Real-time data of SKY module
  • User-friendly verification software
  • Works with E-Val Pro thermocouple system

Software

ValSuite software is the industry's leading verification and calibration software and is highly credited with countless reporting options, impeccable data integrity, user calibration options and data analysis capabilities.

 

The software is used to run all of ELLAB's hardware and can combine various measurement systems into a unified study. In addition to its wide range of features, ValSuite complies with FDA 21 CFR Part 11 and is designed according to the GAMP guidelines. In addition, the software automatically generates PDF reports with clear pass/fail indications and can be run on stand-alone PCs or networked computers.

✓ Complies with FDA 21 CFR Part 11
✓ safe and reliable
✓  Windows Security Options
✓ Compatible with Windows 10
✓ Available in multiple languages
✓ Suitable for various applications
✓ Multiple Reporting Options

  • One software platform for all Ellab devices
  • Combine TrackSense data logger and E-Val Pro wired thermocouple system in the same session
  • Automatic PDF report generation with clear pass/fail indication
  • Runs on standalone PC or network

Applications

環境試驗箱

Environmental test chambers are suitable for a variety of applications: product shelf-life, stability and packaging testing, light and temperature assessment studies, electronic component aging, plant growth, and more.

They are characterized by precise control of parameters such as temperature and relative humidity.

cold store

Refrigeration is used to prevent further biological activity, such as spoilage of food and medical products.The key parameter for the storage period is temperature, but often humidity is also monitored to understand drying conditions.

蒸氣滅菌

製藥和醫療行業的高壓滅菌器必須符合既定的標準和規範(EN 285 和 ISO 17665),以確保其製程持續提供安全和無菌的結果。

為此,高壓滅菌器需要定期進行鑑定和驗證,以證明它們可以在合格的參數範圍內進行滅菌。

Hydrogen Peroxide

Hydrogen peroxide sterilization is a low-pressure, low-temperature, non-toxic process used in the pharmaceutical and medical industries to reduce the level of infectious agents. This process is preferred for products that cannot withstand the heat of a typical high-pressure sterilizer or have space for diffusion limitations, such as low-lumen medical equipment.

Key parameters include pressure/vacuum, transmitted RF energy, H2O2 concentration and temperature. Humidity can also be measured.

Related Products

Contact Hongke to help you solve your problems.