GRD1R LH Temperature Gradient Incubator
Easily realize multi-sample and multi-hole temperature difference test.
Applications
Seed Dormancy
Seeds germinate at constant or variable temperatures.
Germination Rate Measurement
Other applications (seed protection, biofuels, food crop research)
GRD1R LH Temperature Gradient Incubator
The GRD1R LH Temperature Gradient Culture Plate is designed to provide easy access to feedback results of seeds, small plants and animals, microorganisms, and various small components or materials at different temperatures. The design principle is to heat one side of the aluminum plate while cooling the other side, thus creating a temperature gradient on the plate surface, and rotating the plate through the gradient again by 90 degrees to create a porous temperature difference result.
The gradient system is capable of multi-directional temperature cycling. The system can trace the temperature gradient in a defined direction within 24 hours, after which the system automatically rotates the direction of the gradient by 90 degrees in the layout, and then cycling back to the original position in order to produce orthogonal data between the highest and lowest temperatures.
Advantageous highlights of temperature gradient culture plates
Real-time temperature control
The temperature gradient can be set to flip the gradient 90 degrees at any time to create a temperature difference test. 24-hour adjustable timer allows 2-stage control/switching of the gradient direction over a 24-hour cycle.
Multi-sample testing
Plexiglass panels can be moved for warming, and are divided into 196 incubation zones, each with a different temperature, allowing a wide range of products to be tested.
Safeguarding Security Experiments
Over-temperature protection, each edge is protected by a fixed temperature disconnect.
Durable and rugged all-in-one system with caster mounts.
Accurate Data Collection and Analysis
Multi-channel Squirrel data acquisition records temperature and time; 5 sensors are located at the 4 corners and in the middle for analysis by PC.
Model Selection
Dimensions h×d×w (mm) | 1030 x 1020 x 1020 |
Temperature Range Cold Side Heating | +5°C to 30°C Room temperature +5°C to 50°C |
Stability ℃ | ±0.5 |
Display Method | Digital |
Display Resolution ℃ | 1 |
Built-in temperature/time recording sensor | Standard |
Working area mm | 760 x 760 |
Power 230V W | 2050 |
EMC (radiation standard) | Class A |
Weight kg | 43 |